CGM Measurement data - reduced FHIR
Case 1 - With valid CGM, trend and quality values
Sample Scenario
| Data Point | Value | unit |
|---|---|---|
| CGM UUID | BBBB | |
| SessionId | AAAA | |
| CGM Glucose Concentration | 118 | mg/dL |
| CGM Trend Information | -0.5 | mg/dL/min |
| CGM Quality | 90 | % |
| Time Offset | 200 | min |
| Time Stamp | 2020-08-04T15:00:16Z | UTC |
| Flags | 32 - BatteryEmpty | When more than 1 flag is set the sum of those integers is sent. Possible values are LO = 1; HI = 2 ; TempHI = 4 ; TempLO = 8 ; DeviceFault = 16 ; BatteryEmpty = 32 ; SensorMalfunction = 64 ; CalibrationRequired = 128 ; RateOfChangeMin = 256 ; RateOfChangeMax = 512 ; InvalidValue = 1024 ; TypeIncorrectForDevice = 2048 ; DeviceSpecificAlert = 4096 ; CalibrationNotAllowed = 8192 ; CalibrationRecommended = 16384 ; SessionStopped = 32768 ; SensorRateDecreaseExceeded = 65536 ; SensorRateIncreaseExceeded = 131072 ; |
Note: Comments are provided in green in the JSON below
{
"resourceType": "Observation", //FHIR resource type
"meta": { //Should be sent by Confidence saying which profile the below resource conforms to. It is always the same for a DeviceComponent resource
"profile": [
"http://roche.com/fhir/rdc/StructureDefinition/vitalsigns-observation-reduced"
]
},
"identifier": [
{
"system": "http://roche.com/fhir/rdc/sid/observation/confidence",//always same for resourceType Observation
"value": "BBBB" //CGM UUID - unique-identifier-for-the-cgm-reading
}
],
"status": "final", //always final for CGM
"code": {
"text": "CGMGlucoseConcentration" //static text describing that the resource is for a CGM measurement
},
"effectiveDateTime": "2020-08-04T15:00:16Z", //Timestamp of the measurement in zulu/UTC
"valueQuantity": {
"value": 118, //CGM value in mg/dL
"unit": "mg/dL"
},
"component": [
{
"code": {
"text": "CGMTrendInformation" //static text describing that the value is for a CGM trend/rate of change
},
"valueQuantity": {
"value": -0.5, //CGM trend in mg/dL/min
"unit": "mg/dL/min"
}
},
{
"code": {
"text": "CGMQuality" //static text describing that the value is for CGM quality
},
"valueQuantity": {
"value": 90, //CGM quality in %
"unit": "%"
}
}
],
"related": [
{
"type": "derived-from",
"target": {
"display": "AAAA" //SessionID - unique identfier for the session
}
}
],
"extension": [
{
"url": "http://roche.com/fhir/rdc/StructureDefinition/time-offset",
"valueQuantity": {
"value": 200, //Time offset from the start of the session in minutes
"unit": "min"
}
},
{
"url": "http://roche.com/fhir/rdc/StructureDefinition/patient-timezone",
"valueString": "+02:00" //timezone of the patient when the CGM reading was taken. ** Refer timezone extension page for possible combination of timezone displayed.
},
{
"url": "http://roche.com/fhir/rdc/StructureDefinition/flags",
"valueString": "32" //sum of flags
}
]
}
}
** Refer timezone extension page for possible combination of timezone displayed HERE
Case 2 - With NaN for CGM value and NULL for trend and quality
Sample Scenario
| Data Point | Value | unit |
|---|---|---|
| CGM UUID | CCCC | |
| SessionId | AAAA | |
| CGM Glucose Concentration | NaN | |
| CGM Trend Information | NULL | |
| CGM Quality | NULL | |
| Time Offset | 200 | min |
| Time Stamp | 2020-08-04T15:15:16Z | UTC |
| Flags | 160 ( 32 - BatteryEmpty ; 128 - CalibrationRequired ) | When more than 1 flag is set the sum of those integers is sent. Possible values are LO = 1; HI = 2 ; TempHI = 4 ; TempLO = 8 ; DeviceFault = 16 ; BatteryEmpty = 32 ; SensorMalfunction = 64 ; CalibrationRequired = 128 ; RateOfChangeMin = 256 ; RateOfChangeMax = 512 ; InvalidValue = 1024 ; TypeIncorrectForDevice = 2048 ; DeviceSpecificAlert = 4096 ; CalibrationNotAllowed = 8192 ; CalibrationRecommended = 16384 ; SessionStopped = 32768 ; SensorRateDecreaseExceeded = 65536 ; SensorRateIncreaseExceeded = 131072 ; |
Note: Comments are provided in green in the JSON below
{
"resourceType": "Observation", //FHIR resource type
"meta": { //Should be sent by Confidence saying which profile the below resource conforms to. It is always the same for a DeviceComponent resource
"profile": [
"http://roche.com/fhir/rdc/StructureDefinition/vitalsigns-observation-reduced"
]
},
"identifier": [
{
"system": "http://roche.com/fhir/rdc/sid/observation/confidence", //always same for resourceType Observatio
"value": "CCCC" //CGM UUID - unique-identifier-for-the-cgm-reading
}
],
"status": "final", //always final for CGM
"code": {
"text": "CGMGlucoseConcentration" //static text describing that the resource is for a CGM measurement
},
"effectiveDateTime": "2020-08-04T15:15:16Z", //Timestamp of the measurement in zulu/UTC
"dataAbsentReason": {
"text": "NaN" //text describing that the value of CGM is NaN
},
"related": [
{
"type": "derived-from",
"target": {
"display": "AAAA" //SessionID - unique identfier for the session to which this CGM reading belongs
}
}
],
"extension": [
{
"url": "http://roche.com/fhir/rdc/StructureDefinition/time-offset",
"valueQuantity": {
"value": 200, //Time offset from the start of the session in minutes
"unit": "min"
}
},
{
"url": "http://roche.com/fhir/rdc/StructureDefinition/patient-timezone",
"valueString": "+02:00" //timezone of the patient when the CGM reading was taken ** Refer timezone extension page for possible combination of timezone displayed
},
{
"url": "http://roche.com/fhir/rdc/StructureDefinition/flags",
"valueString": "160" //sum of flags
}
]
}
** Refer timezone extension page for possible combination of timezone displayed HERE