EMWR - expanded FHIR
Sample Scenario
| Data Point | Value | unit | Comment |
|---|---|---|---|
| SessionId | CGMSesssion AAAA | ||
| InstanceId | EEMMWW | mg/dL | |
| MessageSource | 2 | Possible values are 0 = None, 1 = Application, 2 = CGM | |
| EventName | M_CALIBRATION_REQUIRED | String | Possible values are M_CALIBRATION_REQUIRED; M_CGM_BATTERY_LOW; M_CGM_SENSOR_MALFUNCTION_WITH_SESSION_STOP; W_CGM_SENSOR_MALFUNCTION_WITHOUT_SESSION_STOP; R_CGM_SESSION_END; M_END_OF_OPERATION; M_CGM_CONNECTION_LOST; M_SENSOR_TEMP_LOW; M_SENSOR_TEMP_HIGH; M_RC_COMM_FAILED; W_CGM_CALIBRATION_OUT_OF_RANGE; M_HYPO_MESSAGE; W_HYPER_MESSAGE; E_DATA_INTEGRITY_ERROR; M_RC_COMM_FAILED; W_HYPO_PREDICTION_MESSAGE; W_LENGTH_OF_STAY_MESSAGE; M_CALIBRATION_CRITICAL |
| TimeOffsetOccurance | 200 | min | |
| State | Snoozed | String | Possible values are NotProcessed, Processed, Snoozed, Dismissed |
| SnoozeWakeupTime | 2020-08-05T15:10:16Z | UTC | |
| TimeWhenProcessed | 2020-08-05T15:15:16Z | UTC | |
| Type | Warning | String | Possible values are Error; Maintenance; Warning; Reminder; Message |
| Time Stamp | 2020-08-04T15:00:16Z | UTC |
EWM Observation resource for Calibration BG measurement
A reduced format of this resource will be sent from Confidence. See the section EMWR - reduced FHIR. The facade needs to further expand that resource by adding the other relavent fields as per our implementation guide. Details are provided in the comments below in green.
Note: Please see the section Detailed modelling for possible values for how the resource needs to be built for different values of EMWR Type, EMWR Event Name and Message Source
Below is an example JSON for the EWM observation resource in expanded FHIR format
{
"resourceType": "Observation",
"id": "qe6f0b94-1519-4f95-b50c-758fa3d4ac55", //To be managed by FHIRSTACK
"meta": {
"profile": [ //To be updated by facade - FHIR profile for Calibration observation for FHIR internal
"http://roche.com/fhir/rdc/StructureDefinition/observation-EWM"
]
},
"identifier": [
{
"system": "http://roche.com/fhir/rdc/sid/observation/confidence",
"value": "EEMMWW" //retain same identifier received from Confidence reduced FHIR
}
],
"status": "final", //retain same status received from confidence reduced FHIR
"category": [ //To be added by facade - category for Calibration observation will always be "Error, Warning, Maintenance, Status"
{
"coding": [
{
"system": "http://roche.com/fhir/rdc/cs",
"code": "EWM",
"display": "Error, Warning, Maintenance, Status"
}
]
}
],
"code": {
"coding": [ //To be added by facade - system, code and display when code.text "Warning" is received in the reduced FHIR.
{
"system": "http://roche.com/fhir/rdc/cs",
"code": "WARNING",
"display": "Warning"
}
],
"text": "Warning" //To be retained as is from the reduced format so that on restore it can be sent back
},
"subject": {
"reference": "Device/0053cfd5-a427-4691-8690-b4d956c6e698" //To be added by facade - Device link
},
"effectiveDateTime": "2020-08-04T15:00:16Z", //Retain same value from Confidence reduced FHIR
"performer": [
{
"reference": "Patient/2P0tI8p-SF9-6850" //To be added by facade - patient link
}
],
"valueCodeableConcept": { //To be added by facade - system, code and display when code.text "M_CALIBRATION_REQUIRED" is received in the reduced FHIR.
"coding": [
{
"system": "http://roche.com/fhir/rdc/cs",
"code": "WARNING_49",
"display": "Calibration Required"
}
],
"text": "M_CALIBRATION_REQUIRED" //To be retained as is from the reduced format so that on restore it can be sent back
},
"related": [
{
"type": "derived-from",
"target": {
"reference": "Observation/0053cfd5-a427-4691-8690-b4d956c6e690", //To be added by facade based on the SessionID (target.display) in the reduced FHIR. Link to the relevant session observation
"display": "AAAA" //To be retained as is from the reduced format so that on restore it can be sent back
}
}
],
"device": {
"reference": "DeviceMetric/id-deviceMetric-CGM", //To be added by facade based on the device.display in the reduced FHIR. link to relavent device metric.
"display": "2" //To be retained as is from the reduced format so that on restore it can be sent back
},
"extension": [
{
"url": "http://roche.com/fhir/rdc/StructureDefinition/time-offset",
"valueQuantity": {
"value": 200, //retain value from reduced FHIR
"unit": "min", //retain value from reduced FHIR
"system": "http://unitsofmeasure.org", //To be added by facade - the system and code
"code": "min"
}
},
{
"url": "http://roche.com/fhir/rdc/StructureDefinition/ewm-state",
"valueString": "Snoozed" //retain value from reduced FHIR
},
{
"url": "http://roche.com/fhir/rdc/StructureDefinition/snooze-wakeup-time",
"valueDateTime": "2020-08-05T15:10:16Z" //retain value from reduced FHIR
},
{
"url": "http://roche.com/fhir/rdc/StructureDefinition/time-when-processed",
"valueDateTime": "2020-08-05T15:15:16Z" //retain value from reduced FHIR
},
{
"url": "http://roche.com/fhir/rdc/StructureDefinition/patient-timezone",
"valueString": "+02:00" //retain value from reduced FHIR
},
{
"url": "http://roche.com/fhir/rdc/StructureDefinition/provenance-channel",
"valueString": "<channel-name>" //To be added by facade - fill channel data
},
{
"url": "http://roche.com/fhir/rdc/StructureDefinition/provenance-source",
"valueString": "<source-name>" //To be added by facade - fill source data
}
]
}