Measurement Status Extension

Canonical URL

http://devices.fhir.org/StructureDefinition/observation-measurement-status

Extension Definition

urlFixed Value
valueCodecodeBinding

Example (XML view)

Example of a SpO2 Numeric Observation

<Observation xmlns="http://hl7.org/fhir">
<id value="632" />
<profile value="http://devices.fhir.org/StructureDefinition/NumericObservation" />
<profile value="http://hl7.org/fhir/StructureDefinition/oxygensat" />
</meta>
<extension url="http://devices.fhir.org/StructureDefinition/observation-measurement-status">
<valueCode value="questionable" />
</extension>
<extension url="http://devices.fhir.org/StructureDefinition/observation-measurement-status">
<valueCode value="msmt-state-in-alarm" />
</extension>
<status value="final" />
<system value="http://hl7.org/fhir/observation-category" />
<code value="vital-signs" />
<display value="Vital Signs" />
</coding>
</category>
<system value="http://loinc.org" />
<code value="59408-5" />
<display value="Oxygen saturation in Arterial blood by Pulse oximetry" />
</coding>
<system value="urn:iso:std:iso:11073:10101" />
<code value="150456" />
<display value="MDC_PULS_OXIM_SAT_O2" />
</coding>
</code>
<reference value="Patient/371" />
</subject>
<effectiveDateTime value="2017-06-02T11:04:48+02:00" />
<value value="88" />
<unit value="%" />
<system value="http://unitsofmeasure.org" />
<code value="%" />
</valueQuantity>
<reference value="DeviceMetric/222" />
</device>
</Observation>

Example (JSON view)

Example of a SpO2 Numeric Observation

{
"resourceType": "Observation",
"id": "632",
"meta": {
"profile": [
"http://devices.fhir.org/StructureDefinition/NumericObservation",
"http://hl7.org/fhir/StructureDefinition/oxygensat"
]
},
{
"url": "http://devices.fhir.org/StructureDefinition/observation-measurement-status",
"valueCode": "questionable"
},
{
"url": "http://devices.fhir.org/StructureDefinition/observation-measurement-status",
"valueCode": "msmt-state-in-alarm"
}
],
"status": "final",
{
"coding": [
{
"system": "http://hl7.org/fhir/observation-category",
"code": "vital-signs",
"display": "Vital Signs"
}
]
}
],
"code": {
"coding": [
{
"system": "http://loinc.org",
"code": "59408-5",
"display": "Oxygen saturation in Arterial blood by Pulse oximetry"
},
{
"system": "urn:iso:std:iso:11073:10101",
"code": "150456",
"display": "MDC_PULS_OXIM_SAT_O2"
}
]
},
"subject": {
"reference": "Patient/371"
},
"effectiveDateTime": "2017-06-02T11:04:48+02:00",
"value": 88,
"unit": "%",
"system": "http://unitsofmeasure.org",
"code": "%"
},
"device": {
"reference": "DeviceMetric/222"
}
}