Measurement Status Extension
Canonical URL
http://devices.fhir.org/StructureDefinition/observation-measurement-status
Extension Definition
observation-measurement-status (Extension) | http://hl7.org/fhir/StructureDefinition/Extension | ||
url | Fixed Value | ||
valueCode | codeBinding |
Example (XML view)
Example of a SpO2 Numeric Observation
<Observation xmlns="http://hl7.org/fhir"> <id value="632" /> <meta> <profile value="http://devices.fhir.org/StructureDefinition/NumericObservation" /> <profile value="http://hl7.org/fhir/StructureDefinition/oxygensat" /> </meta> <extension url="http://devices.fhir.org/StructureDefinition/observation-measurement-status"> <valueCode value="questionable" /> </extension> <extension url="http://devices.fhir.org/StructureDefinition/observation-measurement-status"> <valueCode value="msmt-state-in-alarm" /> </extension> <status value="final" /> <category> <coding> <system value="http://hl7.org/fhir/observation-category" /> <code value="vital-signs" /> <display value="Vital Signs" /> </coding> </category> <code> <coding> <system value="http://loinc.org" /> <code value="59408-5" /> <display value="Oxygen saturation in Arterial blood by Pulse oximetry" /> </coding> <coding> <system value="urn:iso:std:iso:11073:10101" /> <code value="150456" /> <display value="MDC_PULS_OXIM_SAT_O2" /> </coding> </code> <subject> <reference value="Patient/371" /> </subject> <effectiveDateTime value="2017-06-02T11:04:48+02:00" /> <valueQuantity> <value value="88" /> <unit value="%" /> <system value="http://unitsofmeasure.org" /> <code value="%" /> </valueQuantity> <device> <reference value="DeviceMetric/222" /> </device> </Observation>
Example (JSON view)
Example of a SpO2 Numeric Observation
{ "resourceType": "Observation", "id": "632", "meta": { "profile": [ "http://devices.fhir.org/StructureDefinition/NumericObservation", "http://hl7.org/fhir/StructureDefinition/oxygensat" ] }, "extension": [ { "url": "http://devices.fhir.org/StructureDefinition/observation-measurement-status", "valueCode": "questionable" }, { "url": "http://devices.fhir.org/StructureDefinition/observation-measurement-status", "valueCode": "msmt-state-in-alarm" } ], "status": "final", "category": [ { "coding": [ { "system": "http://hl7.org/fhir/observation-category", "code": "vital-signs", "display": "Vital Signs" } ] } ], "code": { "coding": [ { "system": "http://loinc.org", "code": "59408-5", "display": "Oxygen saturation in Arterial blood by Pulse oximetry" }, { "system": "urn:iso:std:iso:11073:10101", "code": "150456", "display": "MDC_PULS_OXIM_SAT_O2" } ] }, "subject": { "reference": "Patient/371" }, "effectiveDateTime": "2017-06-02T11:04:48+02:00", "valueQuantity": { "value": 88, "unit": "%", "system": "http://unitsofmeasure.org", "code": "%" }, "device": { "reference": "DeviceMetric/222" } }